1-point T adjustment

Select 1-point adjustment in the T adjustment menu and follow the steps

below:

1.

Set the probe to a reference temperature. You can follow the
stabilization from the

GRAPH

display. Press

READY

when the

reading is stabilized.

2.

Give the reference temperature by using the arrow buttons and
press

OK

.

3.

To confirm the adjustment, press

YES.

To cancel entering the

adjustment, press

NO

to return to adjustment mode display

without making any changes.

Once confirmed, the adjustment has been carried out. Press

BACK

to

exit the T adjustment mode and return to the main adjustment options.
Press

EXIT

to quit the adjustment mode and return to the basic MI70

view.

2-point T adjustment

Select 2-point adjustment in the T adjustment menu and follow the steps

below:

1.

Set the probe to a lower reference temperature. You can follow the
stabilization from the

GRAPH

display. Press

READY

when the

reading is stabilized.

2.

Give the lower reference temperature by using the arrow buttons
and press

OK

.

3.

Set the probe to a higher reference temperature. You can follow the
stabilization from the

GRAPH

display. Press

READY

when the

reading is stabilized.

4.

Give the higher reference temperature by using the arrow buttons
and press

OK.

5.

To confirm the adjustment, press

YES.

To cancel entering the

adjustment, press

NO

to return to adjustment mode display

without making any changes.

Once confirmed, the adjustment has been carried out. Press

BACK

to

exit the T adjustment mode and return to the main adjustment options.
Press

EXIT

to quit the adjustment mode and return to the basic MI70

view.