Point t adjustment – Vaisala HMT310 User Manual
Page 84

User's Guide _______________________________________________________________________
84 ___________________________________________________________________ M210619EN-E
1-point T adjustment
Select 1-point adjustment in the T adjustment menu and follow the steps
below:
1.
Set the probe to a reference temperature. You can follow the
stabilization from the
GRAPH
display. Press
READY
when the
reading is stabilized.
2.
Give the reference temperature by using the arrow buttons and
press
OK
.
3.
To confirm the adjustment, press
YES.
To cancel entering the
adjustment, press
NO
to return to adjustment mode display
without making any changes.
Once confirmed, the adjustment has been carried out. Press
BACK
to
exit the T adjustment mode and return to the main adjustment options.
Press
EXIT
to quit the adjustment mode and return to the basic MI70
view.
2-point T adjustment
Select 2-point adjustment in the T adjustment menu and follow the steps
below:
1.
Set the probe to a lower reference temperature. You can follow the
stabilization from the
GRAPH
display. Press
READY
when the
reading is stabilized.
2.
Give the lower reference temperature by using the arrow buttons
and press
OK
.
3.
Set the probe to a higher reference temperature. You can follow the
stabilization from the
GRAPH
display. Press
READY
when the
reading is stabilized.
4.
Give the higher reference temperature by using the arrow buttons
and press
OK.
5.
To confirm the adjustment, press
YES.
To cancel entering the
adjustment, press
NO
to return to adjustment mode display
without making any changes.
Once confirmed, the adjustment has been carried out. Press
BACK
to
exit the T adjustment mode and return to the main adjustment options.
Press
EXIT
to quit the adjustment mode and return to the basic MI70
view.