Chemical purge in powerup, Chemical purge activated manually, Starting manual chemical purge – Vaisala HMT310 User Manual

Page 60: H 60

background image

User's Guide _______________________________________________________________________

60 ___________________________________________________________________ M210619EN-E

NOTE

To activate the new interval setting immediately, reset the transmitter.

Chemical Purge in Powerup

This command enables or disables the function which turns on chemical

purge within 10 seconds from the powerup of the device.

Syntax: PURR x

Where
x = ON/OFF (default = OFF)

NOTE

When you enable this function, wait about 8 min after powerup before

taking measurements. The powerup chemical purge locks the output

values for the first operation minutes.

Chemical Purge Activated Manually

Chemical purge needs to be performed always before calibration (see

Chapter Calibration and Adjustment on page 69) or whenever there is

reason to believe that the sensor has been exposed to an interfering

chemical. Make sure that the temperature of the sensor has come down to

normal before starting calibration.

Before starting chemical purge note the following:

- The sensor must be protected with a PPS grid with stainless steel

netting, a stainless steel sintered filter or with a membrane SST filter.

- The sensor temperature needs to be below 100 °C. At higher

temperatures the chemicals evaporate spontaneously from the sensor

and the chemical purge is not necessary.

Starting Manual Chemical Purge

Activate chemical purge with the PURGE command.

Syntax: PURGE

Example:

>purge

Purge started, press any key to abort.

>