Chapter 6 ___________________________________________________ Calibration and Adjustment
VAISALA ________________________________________________________________________ 71
Relative Humidity Calibration after
Sensor Change
Syntax: FCRH
The transmitter asks and measures relative humidity and calculates the
calibration coefficients. This two-point adjustment needs to be performed
after a sensor change. Follow the more detailed calibration instructions
on page 73, but instead of the CRH command, use the FCRH command.
Example:
>fcrh
RH
:
1.82 1. ref
?
0
Press any key when ready
RH
: 74.22
2. ref
? 75
OK
>
The
OK
indicates that the calibration has succeeded.
Set Calibration Information Text
Syntax: CTEXT
1.
Remove the transmitter unit from the mounting plate (see
Mounting the Transmitter/Removing the Transmitter Unit on page
22) and press the adjustment button once (see Figure 18 on page
2.
Issue the LI command and set the offset value to 0 and gain value
to 1.
3.
Issue command CTEXT and enter calibration information text.
Example:
>ctext
Cal. info : Vaisala/HEL ? HMK15
>