Chapter 6 ___________________________________________________ Calibration and Adjustment

VAISALA ________________________________________________________________________ 71

Relative Humidity Calibration after

Sensor Change

Syntax: FCRH

The transmitter asks and measures relative humidity and calculates the

calibration coefficients. This two-point adjustment needs to be performed

after a sensor change. Follow the more detailed calibration instructions

on page 73, but instead of the CRH command, use the FCRH command.

Example:

>fcrh

RH

:

1.82 1. ref

?

0

Press any key when ready

RH

: 74.22

2. ref

? 75

OK

>

The

OK

indicates that the calibration has succeeded.

Set Calibration Information Text

Syntax: CTEXT

1.

Remove the transmitter unit from the mounting plate (see

Mounting the Transmitter/Removing the Transmitter Unit on page

22) and press the adjustment button once (see Figure 18 on page

74).

2.

Issue the LI command and set the offset value to 0 and gain value

to 1.

3.

Issue command CTEXT and enter calibration information text.

Example:

>ctext

Cal. info : Vaisala/HEL ? HMK15

>