Setting Humidity Sensor Heating

When HMT310 leaves the factory the sensor heating follows the factory

default values. You can enable or disable the function, change the RH

limit and define the heating temperature and duration of the heating

period.

Sensor Heating ON/OFF

Enables or disables the sensor heating.

Syntax: XHEAT x

Where
x = ON/OFF (default = OFF)*

* In the HMT317 fuel cell model, XHEAT is enabled in the factory

settings.

Example:

>xheat on

Extra heat : ON

>xheat off

Extra heat : OFF

>

Setting Heating Parameters

Whenever the RH value seen by the sensor exceeds the predefined RH

limit the humidity sensor is warmed up to the predefined temperature.

The duration of the heating can also be defined.

After the heating cycle the humidity conditions are checked and a new

sensor heating is activated if the predefined conditions are reached again.

Syntax: XHEAT

Type the values after the question mark. The ranges available:

- Extra heat RH limit: 0 ... 100 %RH (default: 95 %RH, heating

function starts up above the setpoint)

- Extra heating temperature: 0 ... 200 °C (default: 100 °C)
- Extra heating time: 0 ... 255 s (default: 30 s)