Adjusting RH with MI70

Selecting RH in the main adjustment menu brings up a notification about

checking environment settings. Select YES to access the environment

settings before performing the RH adjustment, or NO to continue to RH

adjustment without changing the current environment settings. Select

ADJUST to access the RH adjustment options.
The following RH adjustment options are available:

- 1-point adjustment
- 2-point adjustment
- LiCl – NaCl auto

1-point RH Adjustment

Usually it is recommended to make an adjustment in two reference

humidities. If the adjustment is done by using one reference humidity (1-

point adjustment), select the reference humidity so that it represents the

measuring environment. Make the adjustment in one point as instructed

in the 2-point RH Adjustment instructions (select 1-point adjustment

instead of 2-point adjustment and follow the display instructions).

2-point RH Adjustment

Select 2-point adjustment in the RH adjustment menu and follow the

steps below:

1.

Set the probe to a lower reference relative humidity. You can
follow the stabilization from the

GRAPH

display. Press

READY

when the reading is stabilized.

2.

Give the lower reference humidity value by using the arrow
buttons. Press

OK

.

3.

Set the probe to the higher reference humidity. Press

READY

when the reading is stabilized. You can follow the stabilization
from the

GRAPH

display.

4.

Give the higher reference relative humidity value by using the
arrow buttons. Press

OK

.

5.

To confirm the adjustment, press

YES.

To cancel entering the

adjustment, press

NO

to return to adjustment mode display

without making any changes. Note that if the difference between

the two references is less than 50 %, adjustment cannot be done.

Once confirmed, the adjustment has been carried out. Press

BACK

to

exit the RH adjustment mode and return to the main adjustment options.