Chapter 6 ___________________________________________________ Calibration and Adjustment

VAISALA ________________________________________________________________________ 81

Adjustment Using an MI70 Indicator

You can use the MI70 handheld indicator to adjust the following in

HMT310:

- Temperature (T)
- Relative humidity (RH)
- Pressure compensation value for humidity measurement (environment

settings)

Note that starting the calibration in MI70 requires pressing the

adjustment button on the HMT310 board. See Figure 18 on page 74 for

the location of the adjustment button. To return to normal operation

mode, use the RESET command (see section Reset Transmitter on page

53).

NOTE

When adjusting both relative humidity and temperature, make the

temperature adjustment first.

Starting MI70 Adjustment

1.

Connect the HMT310 series transmitter to the MI70 handheld

indicator using the connection cable (optional accessory, Vaisala

order code DRW216050SP).

2.

Turn on the MI70.

3.

Press the HMT310 adjustment button (see Figure 18 on page 74) to

start the adjustment mode. A notification about starting the

adjustment mode is displayed on the MI70 screen.

4.

Press OK to enter the adjustment menu. The available adjustment

options are displayed (RH or T). You can also view the date of last

adjustment.

Use the arrow buttons to highlight RH or T and press SELECT to choose

the parameter for adjustment.

For RH adjustment instructions, see section Adjusting RH with MI70.

For T adjustment instructions, see section Adjusting T with MI70.