Set Calibration Date
Syntax: CDATE yyyy mm dd
1.
Remove the transmitter unit from the mounting plate (see
Mounting the Transmitter/Removing the Transmitter Unit on page
22) and press the adjustment button once (see Figure 18 on page
2.
Issue the LI command and set the offset value to 0 and gain value
to 1.
3.
Issue the CDATE command and set the calibration date.
Example:
>cdate 2001 12 11
Calibration : 2001-12-11
>
Analog Output Calibration
Syntax: ACAL
Connect the HMT310 to a multimeter in order to measure either current
or voltage depending on the selected output type. Issue the ACAL
command. Then type the multimeter readings and press Enter.
Example (current outputs):
>acal
Ch1
I1 (mA) ?
2.046
Ch1
I2 (mA) ?
18.087
Ch2
I1 (mA) ?
2.036
Ch2
I2 (mA) ?
18.071
>