Set Calibration Date

Syntax: CDATE yyyy mm dd

1.

Remove the transmitter unit from the mounting plate (see

Mounting the Transmitter/Removing the Transmitter Unit on page

22) and press the adjustment button once (see Figure 18 on page

74).

2.

Issue the LI command and set the offset value to 0 and gain value

to 1.

3.

Issue the CDATE command and set the calibration date.

Example:

>cdate 2001 12 11

Calibration : 2001-12-11

>

Analog Output Calibration

Syntax: ACAL

Connect the HMT310 to a multimeter in order to measure either current

or voltage depending on the selected output type. Issue the ACAL

command. Then type the multimeter readings and press Enter.

Example (current outputs):

>acal

Ch1

I1 (mA) ?

2.046

Ch1

I2 (mA) ?

18.087

Ch2

I1 (mA) ?

2.036

Ch2

I2 (mA) ?

18.071

>