Table 8, Other commands, Table 9 – Vaisala HMT310 User Manual

Page 41: Setting, scaling and testing analog outputs, Table 10, Calibration and adjustment, Table 11, Chemical purge, Table 12, Sensor heating

background image

Chapter 4 _________________________________________________________________ Operation

VAISALA ________________________________________________________________________ 41

Table 8

Other Commands

Command

Description

?

Check transmitter settings

??

Check transmitter settings in POLL mode

ECHO [

ON/OFF]

Serial bus echo

FIND

All devices in POLL mode send their

addresses

HELP

List commands

PRES

Set ambient pressure for calculations (non-

volatile)

XPRES

Set ambient pressure for calculations

FILT

Set result filtering

RESET

Reset transmitter

ERRS

Display error messages


Table 9

Setting, Scaling and Testing Analog Outputs

Command

Description

AMODE

Set analog outputs (0/4 ... 20 mA and 0 …

5/10 V, 1 …5 available through scaling)

ASEL

Select parameter for analog outputs

ASCL

Scale analog outputs

ITEST

Test analog outputs

AQTEST

Test analog outputs for desired readings

AERR

Set error outputs


Table 10

Calibration and Adjustment*

Command

Description

CRH

Relative humidity calibration

CT

Temperature calibration

LI

Revert factory calibration

L

View the currently applied standard user

adjustment offset and gain

FCRH

RH calibration after sensor change

CTEXT

Set calibration information text

CDATE

Set calibration date

ACAL

Analog output calibration

MPC

Multipoint RH calibration

*These commands are presented in section Calibration and Adjustment

Commands on page 70.
Table 11

Chemical Purge

Command

Description

PUR [ON/

OFF]

Enable or disable automatic chemical

purge

PURGE

Activate chemical purge

PURR [ON/

OFF]

Enable or disable automatic chemical

purge in power-up


Table 12

Sensor Heating

Command

Description

XHEAT [ON/OFF]

Enable or disable sensor heating