Automatic Chemical Purge Settings

Turn ON/OFF Automatic Chemical Purge

This command enables or disables automatic chemical purge.
Syntax: PUR x

Where
x = ON/OFF


Example:

>pur off

Chemical Purge : OFF

>pur on

Chemical Purge : ON

>

Set Chemical Purge Interval

If the sensor is exposed to chemicals it is recommended to have the

chemical purge done at least once in 720 min (= 12 hours). In

applications where the chemical exposure is not likely, the interval may

be longer.

In normal use, it is not recommended to change other parameters than the

purge interval. If changing any other parameters is required, always take

note of the current settings before making changes so you are able to

restore the configuration.

Syntax: PUR

Issue the PUR command and type the new interval (in minutes). Press

Enter without entering any values to keep the other parameters in their

current configuration and finish the command. The maximum interval is

14400 minutes (= 10 days).

Example:

>pur

Chemical Purge : ON

Interval min : 1440 ? 720

Duration s : 120 ?

Settling s : 240 ?

Power : 160 ?

Temp diff : 0.50 ?

>