Chapter 5 _______________________________________________________________ Maintenance

VAISALA ________________________________________________________________________ 65

CHAPTER 5

MAINTENANCE

This chapter provides information that is needed in basic maintenance of

the HMT310.

Periodic Maintenance

Calibration Interval

HMT310 is fully calibrated and adjusted as shipped from factory. Typical

calibration interval is one year. In certain applications it may be

advisable to do more frequent checks. Calibration needs to be done

always when there is reason to believe that the device is not within the

accuracy specifications.

You can carry out calibration and adjustment by yourself, or you can

send the transmitter to Vaisala Service Centers for recalibration. For

more information on calibration, see section Calibration and Adjustment

on page 69.

Replacing Consumables

Changing the Filter

Replace the filter if it is contaminated. New filters are available from

Vaisala; see section Spare Parts and Accessories on page 93.

CAUTION

Be careful when changing the filter, since it is easy to break the sensor

when the filter is removed.

CAUTION

If the transmitter is installed in a pressurized chamber, always make sure

that the pressure of the chamber is equalized with the ambient pressure

prior to removing the transmitter. When the transmitter is removed for

maintenance, cap the hole with a capped nut. This way, the chamber can

be pressurized although the transmitter is not in place.
Plugs are available from Vaisala for ISO threaded connections.