Calibration and Adjustment Commands

Revert to Factory Calibration

Syntax: LI

This command reverts only the CRH calibration (see calibration

instructions, starting on page 73).

1.

Remove the transmitter unit from the mounting plate (see

Mounting the Transmitter/Removing the Transmitter Unit on page

22) and press the adjustment button once (see Figure 18 on page

74).

2.

Issue the LI command and enter value 0 for offset and value 1 for

gain.

3.

Reset the transmitter by issuing the RESET command. The

transmitter returns to normal mode.

Example:

>li

RH offset : -0.6000000 ? 0

RH gain : 1.00000000 ? 1

T offset : 0.00000000 ? 0

T gain : 0.40000000 ? 1

>

View Current Offset and Gain Adjustment

View the currently applied standard user adjustment offset and gain using

the L command.

Syntax: L

Example:

>l

RH offset : 0.00000000

RH gain : 1.00000000

T offset : 0.00000000

T gain : 1.00000000

>