Temperature Calibration and Adjustment (in
One Point)
1.
Remove the transmitter unit from the mounting plate (see
Mounting the Transmitter/Removing the Transmitter Unit on page
22) and press the adjustment button once (see Figure 18 on page
2.
Remove the probe filter (see instructions on page 65) and insert the
probe into the reference temperature.
3.
Let the sensor stabilize.
4.
Issue the CT command and press Enter.
5.
Type C and press Enter a few times to check if the reading is
stabilized.
6.
When the reading is stabilized, type the
Ref1
reference temperature
after the question mark and press Enter three times.
>ct
T : 16.06 Ref1 ? c
T : 16.06 Ref1 ? c
T : 16.06 Ref1 ? c
T : 16.06 Ref1 ? c
T : 16.06 Ref1 ? c
T : 16.06 Ref1 ? 16.0
Press any key when ready ...
T : 16.06 Ref2 ?
OK
>
OK indicates that the calibration has succeeded.
7.
Take the probe out of the reference conditions and replace the
filter. When using the stainless steel filter, take care to tighten the
filter properly (recommended force 130 Ncm).
8.
If needed, record the calibration information (text and date) to the
memory of the transmitter, see Set Calibration Information Text on
page 71 and Set Calibration Date on page 72.
9.
Reset the transmitter by issuing the RESET command. The
transmitter returns to normal mode.