Automatic/manual chemical purge, Figure 16, Effect of chemical purge process – Vaisala HMT310 User Manual

Page 58

background image

User's Guide _______________________________________________________________________

58 ___________________________________________________________________ M210619EN-E

NOTE

Using purge is not recommended when the transmitter is powered with

USB. For information on HMT310 operating voltages, see section Power

Supply Requirements on page 34.

0507-056

Figure 16

Decrease of Sensor Gain Due to Interfering Chemical

and Effect of Chemical Purge Process

Automatic/Manual Chemical Purge

When HMT310 leaves the factory the automatic chemical purge (if

selected) takes place repeatedly with the time intervals set in the factory.

You can change the interval in which the purge takes place by using

serial commands. This may be necessary if the measuring environment

contains high concentrations of interfering chemicals. The automatic

chemical purge can be also turned off if necessary.

NOTE

It is not recommended to alter any other purge settings than the interval.

Chemical purge

Calibration

values

Measured values

after chemical

exposure

Humidity

O

ut

put

si

gna