Table 13, Quantity abbreviations for form command, Table 14 – Vaisala HMT310 User Manual
Page 46: Modifiers
User's Guide _______________________________________________________________________
46 ___________________________________________________________________ M210619EN-E
Table 13
Quantity Abbreviations for FORM Command
Abbreviation
Quantity
RH
Relative humidity
T
Temperature
TDF
Dewpoint/Frostpoint
TD
Dewpoint
A
Absolute humidity
X
Mixing ratio
TW
Wet bulb temperature
PPM
Humid air volume/dry air volume
PW
Water vapor pressure
PWS
Saturation water vapor pressure
H
Enthalpy
NOTE
Only quantities selected when ordering the device can be selected for an
output quantity.
Table 14
Modifiers
Modifier
Description
x.y
Length modifier (whole numbers and decimal places)
#t
Tabulator
#r
Carriage return
#n
Line-feed
""
String-constant
U5
Unit field and length (unit can be changed with command UNIT)
Examples:
>form "TD=" 5.2 TD #r#n
TD= -3.65
>form "TD=" TD U3 #t "TDF=" TDF U3 #r#n
TD= -4.0'C TDF= -3.6'C
>
The command FORM / returns the default output form:
>form /
>send
RH= 28.0 %RH T= 23.3 'C