beautypg.com

Scale analog outputs, Table 17 – Vaisala HMT310 User Manual

Page 55

background image

Chapter 4 _________________________________________________________________ Operation

VAISALA ________________________________________________________________________ 55

Table 17

Output Quantities and their Metric and Nonmetric

Units

Quantity

Metric Unit

Nonmetric Unit

RH Relative Humidity

%RH

%RH

T Temperature

°

C

°

F

TDF Dewpoint/frostpoint

°

C

°

F

TD Dewpoint

°

C

°

F

A Absolute humidity

g/m

3

gr/ft

3

X Mixing ratio

g/kg

gr/lb

TW Wet bulb temperature

°

C

°

F

PPM Humid air volume/dry air

volume

ppm

ppm

PW Water vapor pressure

hPa

psi

PWS Saturation water vapor

pressure

hPa

psi

H Enthalpy

kJ/kg

Btu/lb

NOTE

Only the quantities selected when ordering the device can be selected for

an output quantity.

Examples:

>asel td tdf

Ch1 Td lo : -40.00 'C ?

Ch1 Td hi : 100.00 'C ?

Ch2 Tdf lo : -40.00 'C ?

Ch2 Tdf hi : 60.00 'C ?

>asel x td

Ch1 x lo : 0.00 g/kg ?

Ch1 x hi : 160.00 g/kg ?

Ch2 Td lo : -40.00 'C ?

Ch2 Td hi : 60.00 'C ?

>

Scale Analog Outputs

Syntax: ASCL

Example:

>ascl

Ch1 Td lo : -40.00 'C ?

Ch1 Td hi : 100.00 'C ?

Ch2 x lo : 0.00 g/kg ?

Ch2 x hi : 500.00 g/kg ?