beautypg.com

Changing the sensor – Vaisala HMT310 User Manual

Page 66

background image

User's Guide _______________________________________________________________________

66 ___________________________________________________________________ M210619EN-E

When replacing the filter, wear clean gloves to avoid depositing dirt or

oil on the filter.

Replace the filter as follows:

1.

Turn the filter counterclockwise until it is loose.

2.

Pull the filter straight out carefully. Do not damage the sensor.

3.

Take the new filter, and insert it to the filter thread.

4.

Tighten the new filter to 5 Nm by turning it clockwise.

Changing the Sensor

NOTE

When replacing the sensor, the new sensor must be of the same type as

the old sensor (for example, HUMICAP180R). The sensor type can only

be changed at a Vaisala Service Center. Information on Vaisala sensors

is available online at

www.vaisala.com/sensorinfo

.

Changing HUMICAP180 and HUMICAP180R humidity sensors is

simple. If the transmitter has the chemical purge and/or warmed probe

option which utilizes the HUMICAP180C or HUMICAP180RC sensor,

the change is more difficult as soldering is required to attach the pins of

the integrated temperature sensor element. Vaisala recommends that you

leave the replacement of HUMICAP180C and HUMICAP180RC sensors

to Vaisala Service Centers, but spare parts are available if you want do it

yourself. Soldering instructions are not provided.

Changing the sensor should be considered corrective maintenance, and it

is not necessary in normal operation. If the accuracy of the transmitter

does not seem to be within specification, it is more likely that the

transmitter is in need of calibration and adjustment, and not sensor

replacement. Refer to chapter Maintenance on page 65.

1.

Remove the filter from the probe. See the instructions in section

Changing the Filter on page 65.

2.

Remove the damaged sensor and insert a new one. Handle the new

sensor by the plastic socket. DO NOT TOUCH THE SENSOR

PLATE.

3.

Perform a calibration and adjustment as instructed in section

Relative Humidity Calibration after Sensor Change on page 71.

4.

Attach a new filter on the probe. When using the stainless steel

filter, take care to tighten the filter properly (recommended force

5 Nm).