Measurement output, Start continuous outputting, Stop continuous outputting – Vaisala HMT310 User Manual
Page 42: Set continuous output interval for run mode

User's Guide _______________________________________________________________________
42 ___________________________________________________________________ M210619EN-E
Measurement Output
Start Continuous Outputting
Syntax: R
Starts output of measurements to the peripheral devices (RUN mode); the
only command that can be used is S (stop).
The output mode can be changed with the FORM command.
Example:
>r
RH= 28.0 %RH T= 23.3 'C
RH= 28.0 %RH T= 23.3 'C
RH= 28.0 %RH T= 23.3 'C
RH= 28.0 %RH T= 23.3 'C
RH= 28.0 %RH T= 23.3 'C
RH= 28.0 %RH T= 23.3 'C
Stop Continuous Outputting
Syntax: S
Stops the continuous output. Also the Esc key (on the computer
keyboard) can be used to stop outputting.
Set Continuous Output Interval for RUN
Mode
Syntax: INTV xxx yyy
Where
xxx = Output interval (0 ... 255)
yyy = Unit (s, min or h)
Example:
>intv 1
Output interval: 1 S
>intv 1 min
Output interval: 1 MIN
>intv 1 h
Output interval: 1 H