To start, remove the filter from the probe and insert the probe into a

measurement hole of the driest reference chamber you intend to

use. Proceed from low to high humidity. Remember to allow at

least 30 minutes for the sensor to stabilize at each point.

7.

Once you have completed the list, remove the probe from the last

measurement chamber and re-attach the filter.

8.

Using the serial connection command line, issue the MPC RH

INIT command and start entering the readings and reference points

from the list you prepared in step 6. When you have no more points

to enter, press enter to save the points.

>mpc rh init

1 Reading : ? 12.2

1 Reference: ? 11

2 Reading : ? 34

2 Reference: ? 33

3 Reading : ? 75.5

3 Reference: ? 75

4 Reading : ? 97.3

4 Reference: ? 97

5 Reading : ?

Mpc points saved.

9.

Issue the MPC RH LIST command to verify the list you entered:

>mpc rh list

MPC : OFF

# Reading Reference Correction

1 12.20 11.00 -1.20

2 34.00 33.00 -1.00

3 75.50 75.00 -0.50

4 97.30 97.00 -0.30

10. If you are satisfied with the correction list, enable multipoint

correction by issuing the command MPC RH ON.

>mpc rh on

11. Multipoint adjustment is now complete. To exit adjustment mode

and return to the normal mode, issue the RESET command.

The multipoint adjustment can be disabled with the MPC RH OFF

command, and the correction list can be cleared with the MPC RH

CLEAR command. Note that using MPC configuration commands

requires pressing the adjustment button. The current status can be viewed

with the MPC RH LIST command also outside of the adjustment mode.