CLNT0300.DAT configuration file excerpt:
Note that there is an option to select the desired standard on-line. Due to custody transfer regulations
that differ per country, it is also possible to block that possibility.
10.3.1 F9, F8 Choose standard volume calculation standard
These options and values can also be inputted by Modbus
Note that this window control (and Modbus control) can be blocked from normal use, as described in
the previous paragraph.
15
Can be used to change to the old ASTM-IP (table 53/54, D1250) or LPG standard
volume correction as opposed to the (default) API (D2540, 54C) standard.
NOTE: For both D2540 and D1250 the API 11.2.1M equation is used to calculate
the compressability correction.
15.1 Standard_control c=#0 // 0=API (D2540, Table 54C)
// 1=ASTM-IP (D1250, 1953)
// 2=LPG (GPA)
// 3=ULHC (Unstable Liquid Hydro Carbons)
// with online changeable standards BLOCKED:
// 10=as 0 but blocked
// 11=as 1 but blocked
// 12=as 2 but blocked
// 13=as 3 but blocked