ALTOSONIC V

Operating Manual rev02 E

7.30850.34.00

Page 94 of 106

CLNT0300.DAT configuration file excerpt:

Note that there is an option to select the desired standard on-line. Due to custody transfer regulations

that differ per country, it is also possible to block that possibility.

10.3.1 F9, F8 Choose standard volume calculation standard

These options and values can also be inputted by Modbus

Note that this window control (and Modbus control) can be blocked from normal use, as described in

the previous paragraph.

15

Can be used to change to the old ASTM-IP (table 53/54, D1250) or LPG standard

volume correction as opposed to the (default) API (D2540, 54C) standard.

NOTE: For both D2540 and D1250 the API 11.2.1M equation is used to calculate

the compressability correction.

15.1 Standard_control c=#0 // 0=API (D2540, Table 54C)

// 1=ASTM-IP (D1250, 1953)

// 2=LPG (GPA)

// 3=ULHC (Unstable Liquid Hydro Carbons)

// with online changeable standards BLOCKED:

// 10=as 0 but blocked

// 11=as 1 but blocked

// 12=as 2 but blocked

// 13=as 3 but blocked