KROHNE ALTOSONIC V EN User Manual
Page 20
ALTOSONIC V
Operating Manual rev02 E
7.30850.34.00
Page 20 of 106
REAL PROFILE:
On GENERAL FLOW error “1-4 channels down” the REAL profile is used to correct the channels with
errors. The real profile is sampled at a certain flow rate.
•
The REAL-profile correction has a limited validity.
When the actual profile changes too much, the previously sampled REAL-profile might not be
reliable anymore. The check for profile changes is done through flow-rate difference.
•
When the sampled REAL-profile flow rate differs too much from the actual flow-rate during REAL-
profile correction this is shown as a warning.
CORRECTION WARNINGS shows:
•
If there is too much flow variation for corrections, the corrections go on hold. When the corrections
are on hold the real time profile is used as a standard for correcting the flow.
•
If there are too much flow variations or channels failing, the sampling of the REAL profile goes on
hold. On release the sampling is started at maximum time for sampling a profile.
SYSTEM ERRORS shows:
The status of the system is divided into:
•
System Runtime Warnings. These are caused by system failures. These failures will not influence
the flow measurement.
•
System Runtime Alarms. These are caused by system failures. These failures might influence the
flow measurement.
Identified System Runtime Errors are numbered 1 to 60 are:
4.2.1
SYSTEM ERRORS
Identified System Runtime Errors are numbered 1 to 60, A = alarm, W = warning:
Error no. In function
Problem
Consequence
A : 1
Get RS485 data from
converters
Overrun, missed data
Missed data, message
A : 2
Self test
Error in memory self-test
Non-reliable memory
A : 3
Batch start / stop
Error during saving files of start or stop
File lost but ticket is made
A : 4
Profile correction (REAL)
Error in state_correction
Attempt divide to by zero
W: 5
Read Backup all files
Error in reading backup file
Possible loss of backup file
W: 6
Switching disk
Error in finding a drive
Message
W: 7
System time
A notice that the system time was adjusted
manually or by Modbus.
No consequence for totalisers or
proces time, only on ticket time
A: 8
General program
exectutable certification
CRC of the executable is not correct
executable file is corrupted.
Load a new executable file.
Contact KROHNE service for
help
A: 9
Batch status backup
Status file corrupt
Possible loss of batch status
W: 10
Override values files
Error in opening/closing override value file
Override values not stored but
still in use
A: 11
Batch totaliser backup
Totaliser backup-file corrupt
File lost , message
A: 12
Batch average backup
Average backup-file corrupt
File lost, message
A: 13
Batch ticket create
Error in creating batch ticket file
Ticket itself is made for printing
but lost during saving
W: 14
Opening file (for update)
Error in opening REAL file
File lost, message
W: 15
Closing file (for update)
Error in closing REAL file
File lost, message
W: 16
API settings
Error in file, defaults are loaded and saved
Old settings lost
W: 17
Batch 2
A alarm on batch 2 file (Batch 2 is only
used through Modbus with a Scada
system)
File lost, message
W: 18
Check free disk-space
Error dos_getdiskfree() call
Time-out function 30 s
W: 19
Check free disk-space
Low on disk-space
Time-out function 30 s
W: 20
Ad card overrun
The requested AD card is not noticed
Solve the problem
W: 21
Opening file (for update)
Error opening API table file
File lost, message
W: 22
Value check
1 or more API values defaulted
Check the installed parameters
W: 23
Opening file (for update)
Error opening external flow meter file
File lost, message
W: 24
Value check
Default external flow meter K-factor
Check the installed K-factor
W: 25
Counter input
Unable to read Counter value
Read on next entry
A : 26
Calibration MP103 card
MPCA File corrupt
Install backup