ALTOSONIC V

Operating Manual rev02 E

7.30850.34.00

Page 20 of 106

REAL PROFILE:

On GENERAL FLOW error “1-4 channels down” the REAL profile is used to correct the channels with

errors. The real profile is sampled at a certain flow rate.

The REAL-profile correction has a limited validity.

When the actual profile changes too much, the previously sampled REAL-profile might not be

reliable anymore. The check for profile changes is done through flow-rate difference.

When the sampled REAL-profile flow rate differs too much from the actual flow-rate during REAL-

profile correction this is shown as a warning.

CORRECTION WARNINGS shows:

If there is too much flow variation for corrections, the corrections go on hold. When the corrections

are on hold the real time profile is used as a standard for correcting the flow.

If there are too much flow variations or channels failing, the sampling of the REAL profile goes on

hold. On release the sampling is started at maximum time for sampling a profile.

SYSTEM ERRORS shows:

The status of the system is divided into:

System Runtime Warnings. These are caused by system failures. These failures will not influence

the flow measurement.

System Runtime Alarms. These are caused by system failures. These failures might influence the

flow measurement.

Identified System Runtime Errors are numbered 1 to 60 are:

4.2.1

SYSTEM ERRORS

Identified System Runtime Errors are numbered 1 to 60, A = alarm, W = warning:

Error no. In function

Problem

Consequence

A : 1

Get RS485 data from

converters

Overrun, missed data

Missed data, message

A : 2

Self test

Error in memory self-test

Non-reliable memory

A : 3

Batch start / stop

Error during saving files of start or stop

File lost but ticket is made

A : 4

Profile correction (REAL)

Error in state_correction

Attempt divide to by zero

W: 5

Read Backup all files

Error in reading backup file

Possible loss of backup file

W: 6

Switching disk

Error in finding a drive

Message

W: 7

System time

A notice that the system time was adjusted
manually or by Modbus.

No consequence for totalisers or
proces time, only on ticket time

A: 8

General program

exectutable certification

CRC of the executable is not correct

executable file is corrupted.

Load a new executable file.

Contact KROHNE service for

help

A: 9

Batch status backup

Status file corrupt

Possible loss of batch status

W: 10

Override values files

Error in opening/closing override value file

Override values not stored but

still in use

A: 11

Batch totaliser backup

Totaliser backup-file corrupt

File lost , message

A: 12

Batch average backup

Average backup-file corrupt

File lost, message

A: 13

Batch ticket create

Error in creating batch ticket file

Ticket itself is made for printing

but lost during saving

W: 14

Opening file (for update)

Error in opening REAL file

File lost, message

W: 15

Closing file (for update)

Error in closing REAL file

File lost, message

W: 16

API settings

Error in file, defaults are loaded and saved

Old settings lost

W: 17

Batch 2

A alarm on batch 2 file (Batch 2 is only
used through Modbus with a Scada

system)

File lost, message

W: 18

Check free disk-space

Error dos_getdiskfree() call

Time-out function 30 s

W: 19

Check free disk-space

Low on disk-space

Time-out function 30 s

W: 20

Ad card overrun

The requested AD card is not noticed

Solve the problem

W: 21

Opening file (for update)

Error opening API table file

File lost, message

W: 22

Value check

1 or more API values defaulted

Check the installed parameters

W: 23

Opening file (for update)

Error opening external flow meter file

File lost, message

W: 24

Value check

Default external flow meter K-factor

Check the installed K-factor

W: 25

Counter input

Unable to read Counter value

Read on next entry

A : 26

Calibration MP103 card

MPCA File corrupt

Install backup