ALTOSONIC V

Operating Manual rev02 E

7.30850.34.00

Page 88 of 106

9.2

RS485/422 card

There are two possible generations of RS485 cards

9.2.1

RS485/422 card: AX4285A

The first generation of RS 485 cards used

DIP SWITCH CH1***

: COM 3 Baseaddress ch#1: 3E8

DIP SWITCH CH2***

: COM 4 Baseaddress ch#2: 2E8

JP1***

: COM3 Interrupt IRQ4

JP2***

: COM4 Interrupt IRQ3

JP3***

: COM3 RS 485 mode

JP4***

: COM3 Serial resistors enabled, No jumpers installed

JP5

: COM4 RS 485 mode as a default

JP6

: COM4 Serial resistors not enabled, jumpers installed

***(=KROHNE Altometer setting)
NOTE:
RS485 mode and RS422 mode for COM4 (Modbus) differs in set-up by:

-

Jumper JP5 RS485 or RS422

-

The external wiring for RS422 and RS485

External wiring AX5285A for Modbus:

The resistors of 120 Ohm must be placed At the ALTOSONIC V wiring terminal