8.4
Analog outputs AD card
The AD card has two 0-10V analog outputs.
Resolution is 12 bits, linearity ±½ bit, settling time 30 microseconds. With additional converters the 0-
10V range can be converted into 4-20 mA signals
•
The AD output function can be disabled/enabled in the Initialisation file: HSET0300.ufp section 4
•
The AD output can be configured in the Initialisation file CLNT0300.dat section 7
•
The signals can be checked on value in the service window: IO.
•
Monitoring is also possible by its calibration program (see Manual: ALTOSONIC V UFP
Calibration and Verification I/O)
HSET0300.ufp section 4
CLNT0300.dat
-----------------------------------------------------------------------------
7
7.1 Out2_mode =#4 //0=DIS 1=flow[m3/h] 2=flow15 3=mass[ton/hr]
4=dens[kg/m3]
//5=c_s[m/s] 6=VCF 7=viscosity[10e-6 m2/s]
//8=dens15[kg/m3] 9=Temp[øC] 10=Pres[bar]
7.2 Out2_min_volt =#0 //Minscale U [V], range= 0 - max_volt [V]
7.3 Out2_max_volt =#10 //Maxscale U [V], range= min_volt - 10 [V]
7.4 Out2_min_unit =#610 //Min outputvalue in [unity] choice
7.5 Out2_max_unit =#1075 //Max outputvalue in [unity] choice
7.6 Out2_tau =#10 //Averaging time tau [s]
7.7 Out3_mode =#7 //0=DIS 1=flow[m3/h] 2=flow15 3=mass[ton/hr]
4=dens[kg/m3]
//5=c_s[m/s] 6=VCF 7=viscosity[10e-6 m2/s]
8=dens15[kg/m3]
//9=Temp[øC] 10=Pres[bar]
7.8 Out3_min_volt =#0 //Minscale U [V], range= 0 - max_volt [V]
7.9 Out3_max_volt =#10 //Maxscale U [V], range= min_volt - 10 [V]
7.10 Out3_min_unit =#0 //Min. outputvalue in [unity] choice
7.11 Out3_max_unit =#150 //Max. outputvalue in [unity] choice
7.12 Out3_tau =#60 //Averaging time tau [s]
4.3 AD_curr_out =#0 //Current outputs disable=0, enable=1