ALTOSONIC V

Operating Manual rev02 E

7.30850.34.00

Page 13 of 106

3.2

Reading initialisation files on input range

Each parameter is checked for its input range.

1. If a parameter is out of range, the software switches to fail-mode.

(Only breakable by pin code 1357)

2. In fail mode a system set-up Error Code is given.

The parameter and its input range are printed on screen. If the Modbus communication is active

the set-up Error Code is also available on this output.

3. If there are no problems at start-up, the software checks whether the CRC-checked data files

correspond with the backup file BACK0300.bin.

This backup file also has a CRC-checksum. Only when the data files do not correspond or the

backup checksum gives an error, a new backup file and checksum are made.

3.3

Batch commands for configuration change and program start up

The following batch commands can be used in DOS mode:
No Name

Description

1

MEAS

Start measure program to measure flow

2

AD

Start calibrate-verify AD card IO (AD-812)

3

FR

Start calibrate-verify Frequency card IO (MP103)

4

CLNT

Edit parameters on IO, Spans etc

5

COMS

Edit communication setup (Modbus, Batchprinter etc)

6

SYST

Edit the system file (syst0300.ufs)

7

TICK

Edit ticket (BOL) layout file

8

HSET

Edit parameters on hardware settings of the UFP

9

SECU

Secure all (configuration, programs, operating system)

10

BACKALL

Make a backup to an empty floppy of flow configuration and OS

11

BACKFLOW

Make a backup to an empty floppy of flow configuration only

12

BACKOS

Make a backup to an empty floppy of Operating System OS only

13

BACKZIP

Make a backup to empty floppy (zipped)

14

FLOW

Edit the flow calibration file

15

REYN

Edit the reynolds calibration file

16

SWRL

Edit the swirl calibration file

Note that these files are custody transfer configurations and a password is required to enable the

changes for the measurement program.