Reading initialisation files on input range
Each parameter is checked for its input range.
1. If a parameter is out of range, the software switches to fail-mode.
(Only breakable by pin code 1357)
2. In fail mode a system set-up Error Code is given.
The parameter and its input range are printed on screen. If the Modbus communication is active
the set-up Error Code is also available on this output.
3. If there are no problems at start-up, the software checks whether the CRC-checked data files
correspond with the backup file BACK0300.bin.
This backup file also has a CRC-checksum. Only when the data files do not correspond or the
backup checksum gives an error, a new backup file and checksum are made.
3.3
Batch commands for configuration change and program start up
The following batch commands can be used in DOS mode:
No Name
Description
1
MEAS
Start measure program to measure flow
2
AD
Start calibrate-verify AD card IO (AD-812)
3
FR
Start calibrate-verify Frequency card IO (MP103)
4
CLNT
Edit parameters on IO, Spans etc
5
COMS
Edit communication setup (Modbus, Batchprinter etc)
6
SYST
Edit the system file (syst0300.ufs)
7
TICK
Edit ticket (BOL) layout file
8
HSET
Edit parameters on hardware settings of the UFP
9
SECU
Secure all (configuration, programs, operating system)
10
BACKALL
Make a backup to an empty floppy of flow configuration and OS
11
BACKFLOW
Make a backup to an empty floppy of flow configuration only
12
BACKOS
Make a backup to an empty floppy of Operating System OS only
13
BACKZIP
Make a backup to empty floppy (zipped)
14
FLOW
Edit the flow calibration file
15
REYN
Edit the reynolds calibration file
16
SWRL
Edit the swirl calibration file
Note that these files are custody transfer configurations and a password is required to enable the
changes for the measurement program.