ALTOSONIC V

Operating Manual rev02 E

7.30850.34.00

Page 24 of 106

that the factor is used in the flow calculation. The correction is lineair and depends on meter

construction.

For example correction per 100 Bar difference for:

a certain ALTOSONIC V 6 inch construction is about +0.04%

a certain ALTOSONIC V 10 inch construction is about +0.045%

As this is a linear function the correction at 50 Bar difference would be half the value (0.02%).

This was calculated using the standard: ISO/CD 17089/1

The correction is described in different standards. To be compliant with these different standards

KROHNE as made the choice to put the correction in a general formula:

)

(

100

1

ref

proces

pb

pb

P

P

C

K

+

=

K

bp

: Correction factor used for the pressure expansion

C

bp

: Lineair pressure correction factor [%/bar], depending on the construction and used standard.

P

proces

: Pressure proces[bar]

P

ref

: Pressure reference [bar],

Note that P expansion correction is disabled at that time, or used pressure is low.

STANDARD VOLUME CORRECTIONS shows:

The temperatures, pressures and densities at proces, standard, densitometer and optional

external flow meter conditions in relation to the correction factors Ctl and Cpl

The correction factors Ctl (temperature correction to 15°C) and Cpl (pressure correction to

1.01325 bar, or 0 barg)

See chapter 5 for more information on the Standard Volume correction