beautypg.com

KROHNE ALTOSONIC V EN User Manual

Page 93

background image

ALTOSONIC V

Operating Manual rev02 E

7.30850.34.00

Page 93 of 106

proces temperature and pressure can be copied to the external temperature and pressure but note

that 1°C difference causes about 0.1% error, and 1 bar difference causes about 0.01% error.

Practice shows that both repeatability and linearity improve when comparing calculated standard

values.

It is possible to compare the proces volumetric total of the external flow meter with the proces

volumetric total of the UFP-V but then the ALTOSONIC V must also be set as only calculating proces

volume.

Description of the controls in this window:

Function keys do the controls of this window, therefore it is only possible to go back to the Main

window.

F1

: Go back to Main window

F2 (or ENTER)

: Disable/enable value change of the K factor manually

F3 (or arrow up)

: If value change is enabled (F2) increase value

F4 (or arrow down)

: If value change is enabled (F2) decrease value

F5 (or )

: If value change is enabled F2) increase step value of change (F3,F4)

F6 (or ) : If value change is enabled (F2) decrease step value of change (F3,F4)

F7 (or )

: Start proving, reset of totals and errors on both UFP-V and External

F8 (or )

: Install found NEW K factor and start prove as described in F7

F10 (or < B> )

: Save configuration if K factor is manually installed

Note: Starting a prove involves resetting of resetable totalisers and occurred alarms.



10.2 Base Sediment and water (BSW)
The Base Sediment and Water (BSW) is inputted by AD input or Modbus as a percentage of the

volume flow.

The actual value of the BS&W percentage is provided through modbus: F7591

or alternatively on an AD input channel if the modbus connection times out for more than 30 seconds.

It is off course also possible to input the value only by the AD-input.

The nett totalisers are calculated by subtracting the BSW percentage from the gross flow (proces,

standard, mass) and totalising this into separate totalisers.



10.3 Other Standard Volume standards than API2540


Other then API2540 volume correction standards are implemented.

ASTM-IP (D1250, 1953)

API 11.2.1M can be used to calculate the compressibility correction.

LPG (GPA) TP25

API 11.2.2M is used to calculate the compressibility correction.

ULHC (Unstable Liquid Hydro Carbons), specially developed for the Russian

Depending on the standard choosen, under F9, F2 the configuration can be made