beautypg.com

KROHNE ALTOSONIC V EN User Manual

Page 42

background image

ALTOSONIC V

Operating Manual rev02 E

7.30850.34.00

Page 42 of 106

4.9.2

Service menu: F3 UFC errors window

Under normal circumstances it is not necessary to view this window.

All data shown here is also available in more common used windows in perhaps other formats or

condensed into less variables.

The status is shown as counters per channel.
There is no history in the counters so previous occurred errors will turn to zero.

Communication errors per communication message (=per channel request):

Parity errors

Error in message length

Wrong start bytes

Framing error UART

Communication status sublimated from communication errors per channel:

Channel state = 0: no errors (status normally)

Channel state = 1: error resulting in a single communication failure (COMFA)

Channel state = 2: comm. failures in succession resulting in a communication alarm (COMMU)

Communication status regarding data skipped or already handled:

Old data

: Counter for data, already handled(Note: normally toggles between 0 and 1).

Overrun

: Counter for data, skipped because of system time shortage (note: cumulative!).