KROHNE ALTOSONIC V EN User Manual
Page 66
ALTOSONIC V
Operating Manual rev02 E
7.30850.34.00
Page 66 of 106
3
At all flowing conditions
No
Yes (continuous pipe
line measurement)
4
At all flowing conditions
No
Yes (continuous pipe
line measurement)
BATCHING_ON 1 and 2 have the following restrictions during a batch:
-No reset possible of resetable totalisers
-No reset of error times but the possibility to reset occurred error messages
•
The previous number of tickets saved is set with MAX_TICKETS. Default is 100 tickets.
Be careful with increasing the number of tickets. Not enough disk space means losing tickets
•
For Continuous Pipe Line Measurement the ticket automatically is printed starting from
HOUR_START
•
For Continuous Pipe Line Measurement the ticket automatically is printed every
HOUR_INTERVAL, but if interval 0 is installed than tickets are only printed on demand
•
By MOD_BUS_CONTROL it is possible to enable the controls through Modbus for batching:
-Start batch
-Stop batch
-Reset printing
-Confirm printing
Or in case of using continuous Pipe Line measurement
-Ticket on demand with reset of values
-Ticket on demand without reset of values
- Reset printing
6.5
Batch status
Batch status
(status window text)
As a value on
Modbus
Explanation
NON
0
No batch active, ready to set-up
SETUP
1
In set-up mode. After set-up is done, it is possible to
start a batch
RUNNING
2
Batch is started
END-BATCH
3
Batch is stopped and ticket is made, then attempt to
END_PRINT
END-PRINT
5
Status during successful printing
END-FAIL
6
If printing fails or printer is busy too long
CONFIRM
7
After successful print job waiting for manual
confirmation
RESET
10
Waiting for reset command after END_FAIL