8 OUTPUT
The output consists of:
•
Frequency output MP103 card
•
Analog output MP103 card
•
Relay outputs MP103 card
•
Analog outputs AD card
•
Digital outputs AD card
•
Modbus communication
8.1
Frequency output MP103 card
Frequency output:
-
Maximum output range is software adjustable 1 – 2000 Hz
-
12V/24V / open-collector selectable by card jumpers
-
There is one output value but there are two physical outputs, these can be phase-shifted 90°/180°
selectable by card jumper to simulate a turbine output for pulse fidelity and integrity check.
The resolution of the frequency output is max 0.016% of the output value. The resolution mentioned is
for a static output value. In practice, the output resolution will be averaged because of the variations in
signal. Over a period of time with different output values the resolution will not be an issue.
The most likely frequency output is the proces volumetric flow (default).
•
The frequency output function can be disabled/enabled in the Initialisation file: HSET0300.ufp
section 3
•
The frequency output can be configured in the Initialisation file: CLNT0300.dat section 5
•
The signals can be checked on value in the service window: IO.
•
Monitoring is also possible by its calibration program (see Manual: ALTOSONIC V UFP
Calibration and Verification I/O)
HSET0300.ufp section 3
CLNT0300.dat section 5
-----------------------------------------------------------------------------
5
5.1 Freq_max =#1000 //Max.scale [Hz], range= 1 - 2000 [Hz]
5.2 Freq_mode =#1 //0=DIS 1=flow[m3/h] 2=flow15 3=mass[ton/hr]
//4=dens[kg/m3] 5=c_s[m/s] 6=VCF 7=viscosity[10e-6
m2/s]
//8=dens15[kg/m3] 9=Temp[øC] 10=Pres[bar]
5.3 Freq_min_unit =#0 //Min outputvalue in [unity]
5.4 Freq_max_unit =#1800 //Max outputvalue in [unity]
5.5 Freq_tau =#0 //Averaging time tau[s]
5.6 Freq_dir_flow =#1 //Directionflow for output frequency: 0=+, 1=+-
3.1 MP_freq_out =#0 //Frequency output 0=disable, 1=enable