ALTOSONIC V

Operating Manual rev02 E

7.30850.34.00

Page 21 of 106

A : 27

Calibration AD card

File corrupt

Install backup

A : 28

Calibration data Densito

Cells

File corrupt

Automatic install of default values

Set the correct values on-line

A : 29

Batch ticket currently
saved

A Requested batch ticket not available for
printing

A ticket by that name was not
saved or had a previous save

error

A : 30

Batch ticket

CRC error in a Batch ticket

A ticket was not saved correctly
or was changed manually

W: 31

Read batch ticket

previously saved

A Requested batch ticket not available for

printing

A ticket by that name was not

saved or had a previous save

error

W : 32

Batch ticket close file

Error in closing a ticket file

Ticket file not closed , probably
because it could not be opened


See for the communication runtime errors also the ALTOSONIC V ModBus Manual.

Err no.

In function

Problem

Consequence

W: 33

Modbus master

Poll block not send due to transmit error

W: 34

Modbus master

Poll block response time-out occurred

W: 35

Modbus master

Invalid Slave ID in response

W: 36

Modbus master

Invalid function in response

W: 37

Modbus master

Response not correct

W: 38

Modbus master

Error handling function 1,2

W: 39

Modbus master

Error handling function 3,4

W: 40

Modbus master

Error handling function 5

W: 41

Modbus master

Error handling function 6

W: 42

Modbus master

Error handling function 15

W: 43

Modbus master

Error handling function 16

W: 44

Modbus master

Exception received

W: 45

Modbus master

Error unpacking Boolean data

W: 46

Modbus master

Error unpacking integer data

W: 47

Modbus master

Error unpacking long integer data

W: 48

Modbus master

Error unpacking float data

W: 49

Modbus master

Error unpacking double data

W: 50

Modbus master/slave

Error incorrect message length

W: 51

Modbus master/slave

Invalid CRC or LRC received

W: 52

Modbus master/slave

Error receive buffer saturated

W: 53

Modbus master/slave

UART error ( parity, framing, overrun )

W: 54

Modbus master/slave

Transmit buffer not empty for new

transmission

W: 55

Modbus slave

Unsupported function requested

W: 56

Modbus slave

Unsupported register(s) requested

W: 57

Modbus slave

Requested data Level and function

mismatch

W: 58

Modbus slave

Too many data point (registers) requested

W: 59

Modbus slave

Error unpacking received data

W: 60

Modbus slave

Broadcast not allowed

Note: Occurred and disappeared alarms and warnings can be reset in the Control menu: F7 RES-

ERR. It is also possible to reset by digital input signal or Modbus Boolean.