KROHNE ALTOSONIC V EN User Manual
Page 10

ALTOSONIC V
Operating Manual rev02 E
7.30850.34.00
Page 10 of 106
Secondary input
Function
Temperature body
For correction of the expansion of the UFS, resulting in a correction
factor K
b
on the measured flow
Temperature proces*
For standard volume correction
Resulting in a correction factor C
tl 15 to proces
on the measured flow
Temperature proces**
For correction on standard calibration volume (Factory use only).
Function is only applicable when the calibration is not only monitored by
the calibration facility but also, with a digital start/stop signal, by the
UFP. The standard calibration volume is the volume measured at a
standard temperature
Temperature densitometer*
For standard volume correction
Resulting in a correction factor C
tl 15 to densito
on the measured flow
Pressure proces*
For standard volume correction
Resulting in a correction factor C
pl proces
on the measured flow
Pressure densitometer*
For standard volume correction
Resulting in a correction factor C
pl densito
on the measured flow
Densitometer density*
The density measured by the densito meter
Density standard*
The density standard with at predefined standard temperature
* = Optional
**= KROHNE Altometer calibration use only