3.4

Start up: system set-up errors

The system SET-UP ERRORS are caused by an improper initialisation such as data–change etc.

If the UFP-V identifies a system set-up error, it switches to fail-mode.

The fail-mode shows the found error and the elapsed proces error time. The mode can only be

stopped by pin code 1357.

Identified set-up errors are:

Error
No.

Function

Problem

How to solve

1

CRC

Error opening: file(filename) to check on CRC

Try to load backup (CRC-function)

2

CRC

Error closing: file(filename) to check on CRC

Try to load backup (CRC-function)

3

CRC

Error opening: CRC-code file(filename)

Try to load backup (CRC-function)

4

CRC

Error closing: CRC-code file(filename)

Try to load backup (CRC-function)

5

CRC

Error length: CRC-code file(filename)

Make new CRC checksum

6

Common, opening file

Error in path: file(filename) not found

Try to load back-up (CRC-function)

7

Not in use

Not in use in this version

8

Common, read in table File(filename), maximum rows exceeded

Put in less data points

9

Common, closing file

Error read in file(filename)

Try to load backup (CRC-function)

10

Common, closing file

Error write in file(filename)

Try to load backup (CRC-function)

11

Read in profiles

Error in file(filename): a parameter <|0.01|

Try to load backup (CRC-function)

12

Not in use

Not in use in this version

13

Check on serial

numbers

Serial numbers in parameter files do not

correspond

Check the serial number in files

14

Initialising Graph driver Graphics error

Is egavga.bgi file in directory ASV0300?

15

File location

Error in finding disk

Check the file locations in
HSET0300.ufp

16

Frequency set-up

Error in set-up frequency output

Follow instructions on screen

17

Common, read in

parameter

Error in a parameter file, bad up-dating, make

sure that ‘#’ is first

Check your last updated file or load

backup (CRC function)

18

Common, read in
parameter

Error in a parameter file, number too large
(more then x characters)

Check your last updated file or load
backup (CRC function)

19

Factory use only

20

Factory use only

21

Not in use

22

Check location

executable

Error in LOCATION_EXE, proces location is

disk x

Change LOCATION_EXE in

HSET0300.ufp

23

Not in use

24

Check parameters on

range

Out of range in file(filename),

parameter(name)=x,
Must be in range x1…x2

Follow the instructions on screen

25

CRC-checksum

outcome

CRC checksum not correct!

Make a new checksum or if not certain

about the data, load the backup (CRC-
function)

26

Not in use

27

CRC-checksum

CRC backup-files checksum not correct

Fill in the correct data in actual files

Backup

28

Batch status files

When the batch mode is enabled and the batch

status files are not found at start-up.

After breaking the fail mode follow the

instruction on screen to insert your last
ticket number

29

Initialisation Printer

When the batch mode is enabled, the printer

software is initialised. On error of initialisation

Check the COMS0300.dat file for errors

in Printer set-up

30

Password

If for any reason the password is lost

Try to load backup (CRC-function)