KROHNE ALTOSONIC V EN User Manual
Page 14

ALTOSONIC V
Operating Manual rev02 E
7.30850.34.00
Page 14 of 106
3.4
Start up: system set-up errors
The system SET-UP ERRORS are caused by an improper initialisation such as data–change etc.
If the UFP-V identifies a system set-up error, it switches to fail-mode.
The fail-mode shows the found error and the elapsed proces error time. The mode can only be
stopped by pin code 1357.
Identified set-up errors are:
Error
No.
Function
Problem
How to solve
1
CRC
Error opening: file(filename) to check on CRC
Try to load backup (CRC-function)
2
CRC
Error closing: file(filename) to check on CRC
Try to load backup (CRC-function)
3
CRC
Error opening: CRC-code file(filename)
Try to load backup (CRC-function)
4
CRC
Error closing: CRC-code file(filename)
Try to load backup (CRC-function)
5
CRC
Error length: CRC-code file(filename)
Make new CRC checksum
6
Common, opening file
Error in path: file(filename) not found
Try to load back-up (CRC-function)
7
Not in use
Not in use in this version
8
Common, read in table File(filename), maximum rows exceeded
Put in less data points
9
Common, closing file
Error read in file(filename)
Try to load backup (CRC-function)
10
Common, closing file
Error write in file(filename)
Try to load backup (CRC-function)
11
Read in profiles
Error in file(filename): a parameter <|0.01|
Try to load backup (CRC-function)
12
Not in use
Not in use in this version
13
Check on serial
numbers
Serial numbers in parameter files do not
correspond
Check the serial number in files
14
Initialising Graph driver Graphics error
Is egavga.bgi file in directory ASV0300?
15
File location
Error in finding disk
Check the file locations in
HSET0300.ufp
16
Frequency set-up
Error in set-up frequency output
Follow instructions on screen
17
Common, read in
parameter
Error in a parameter file, bad up-dating, make
sure that ‘#’ is first
Check your last updated file or load
backup (CRC function)
18
Common, read in
parameter
Error in a parameter file, number too large
(more then x characters)
Check your last updated file or load
backup (CRC function)
19
Factory use only
20
Factory use only
21
Not in use
22
Check location
executable
Error in LOCATION_EXE, proces location is
disk x
Change LOCATION_EXE in
HSET0300.ufp
23
Not in use
24
Check parameters on
range
Out of range in file(filename),
parameter(name)=x,
Must be in range x1…x2
Follow the instructions on screen
25
CRC-checksum
outcome
CRC checksum not correct!
Make a new checksum or if not certain
about the data, load the backup (CRC-
function)
26
Not in use
27
CRC-checksum
CRC backup-files checksum not correct
Fill in the correct data in actual files
Backup
28
Batch status files
When the batch mode is enabled and the batch
status files are not found at start-up.
After breaking the fail mode follow the
instruction on screen to insert your last
ticket number
29
Initialisation Printer
When the batch mode is enabled, the printer
software is initialised. On error of initialisation
Check the COMS0300.dat file for errors
in Printer set-up
30
Password
If for any reason the password is lost
Try to load backup (CRC-function)