7 DATA ACQUISITION

Input data can be divided into:

Data input RS485 card

Digital inputs MP103 card

Frequency inputs MP103 card

Analog inputs AD card

7.1

Data input RS485 card

The data measured by the five converters UFC-V is transferred to the UFP-V by using a half-duplex

Krohne communication protocol based on balanced data transmission (RS485).

The Krohne communication protocol requests the five converters for new measured data. The

incoming data is first checked on parity-errors, framing-errors, and overruns. The data essentially

contains the measured flow from 5 ultrasonic measuring paths, transit time, and error codes. The

converter sends data on every request the UFP makes (about every 35ms).

7.2

Digital inputs MP103 card

The MP103 card has 4 digital inputs.

The digital inputs are normally open (is 0)

The logic level is TTL compatible, maximum 12 VDC.

Channel
no.

Function

Action

0

Reset measured volume, proces-time and error messages Make input ‘1’ to reset

1

Reset error messages

Make input ‘1’ to reset

2

Calibration start-signal (KROHNE Altometer use only)

Make input ‘1’ to arm, make ‘0’ to enable

3

Calibration stop-signal (KROHNE Altometer use only)

Make input ‘1’ to arm, make ‘0’ to enable


The digital input function can be disabled/enabled in the Initialisation files: HSET0300.UFP section

3

The individual channels can be disabled/enabled in the Initialisation files: CLNT0300.dat section 8

The signals can be checked on value in the service window IO.

Monitoring is also possible by its calibration program (see Manual: ALTOSONIC V UFP

Calibration and Verification I/O)

1 2 3 4 5

RS 485 card

in UFP

Requests to UFC

converters 1..5

Reply of UFC converters 1..5
-

Flow percentage –125..+125%

-

Transit time

-

Status overrange

-

Status path failure