beautypg.com

Pass/fail testing – GW Instek GSP-730 User Manual User Manual

Page 61

background image

BASIC OPERATION

59

frequency column of the desired point.

Cursor

4. Enter the new frequency and amplitude of the

point using the keypad and the unit keys.

5. Repeat steps 3-5 for the remaining points (A

maximum of ten points).

6. To delete the selected point, press Delete [F3].

7. To delete all the points, press Delete All… [F4].

The points will revert to their default frequency and
amplitude values.

8. To delete a point from the editing table, press

Delete.

The whole points, including the frequency value and
the amplitude are removed. This command reduces
the number of points used in the limit line.

9. Press Undelete[F5] the restore the last point that

was deleted.

Pass/Fail Testing

Description

The Pass/Fail testing uses the limit lines as
boundaries. When the input signal escapes the
boundary of the limit lines, then the test is
judged as a FAIL, if the signal stays within the
boundary, the test is judged as a PASS.