Determine whether or not the analyzer should capture spectrum

files under a

WARNING

condition.

Note: this may be useful to do so however, if the Warning alarm

conditions are not set correctly then there could be excessive files

created for less meaningful Warning alarm conditions.

An example is a low transmission warning alarm set at 70% for an

application that often runs at less than 70% transmission.

Determine whether or not the analyzer should capture spectrum

files under a

FAULT

condition.

Note: this may be useful to do so however, if the Fault alarm

conditions are not set correctly, then there could be excessive files

created for less meaningful Fault alarm conditions.

These files are often useful for Factory based diagnostics.