<6 BASIC OPERATION> 6-13

IM 11Y01B02-01E-A 4th Edition September 11, 2012-00

ONLINE VALIDATIONS – DO NOT USE FOR

TDLS220

ADVANCED DATA MENU

ALARM HISTORY – shows chronological list of

analyzer’s most recent alarms.

CAL HISTORY – shows chronological list of analyzer’s

recent calibrations.

SPECTRUM CAPTURE – selection of AUTOMATIC

(user will be prompted to select capture interval,

number of UPDATES to trigger capture, RELATIVE

concentration level trigger which is a % of reading

change, or ABSOLUTE concentration level to trigger

capture); in addition the software will prompt for

number of spectra to capture when a Warning or

Fault occurs. MANAUL selection will result in spectra

capture only when requested by user.

RECORD RESULT DATA – The default setting during

normal operation is “User Data”. The system should

only be switched to “Factory Data” when advised by

Yokogawa Laser Analysis Division. Note: recording

Factory Data is only for specific diagnostic purposes

and should not be selected under normal operation.