<6 BASIC OPERATION> 6-15

IM 11Y01B02-01E-A 4th Edition September 11, 2012-00

NON-PROCESS PATH LENGTH

This is the optical path length inside the laser & detect

measurement unit. Typical values would be 3.5” to 4” with the

default value to use at 3.75”. Note that this dimension will vary

between analyzers.

Contact Yokogawa if unsure.

NON-PROCESS PRESSURE

This is the pressure of the non-process gas. Typically,

this will be close to atmospheric pressure of 1.01BarA or

14.7PsiA. Check the actual operating conditions and enter the

appropriate value. Contact Yokogawa if unsure.

NON-PROCESS TEMPERATURE

This is the temperature of the non-process gas with two

modes of input:

ACTIVE AMBIENT - ambient gas temperature derived from

integral sensor on detector circuit with offset adjustment

(typically -18 deg C). Note this off-set is due to the internal

heat of the controller unit, where the detect board is located in

TDLS220

FIXED – manual input of fixed temperature value

NON-PROCESS CONCENTRATION

This is the concentration of the non-process measurement

gas. For Instrument Air purge gas, the typical value would be

20.9% O2.

If measuring CO in the process gas and the purge gas is

Instrument Air, then these parameters are not applicable

because the CO concentrations typically found in Instrument

Air are below practical detection limits.