<6 BASIC OPERATION> 6-8

IM 11Y01B02-01E-A 4th Edition September 11, 2012-00

UNITS – selection of units for path length (in, ft, cm, m),

pressure (psiA, barA, kPa, torr, atm) and temperature (ºF,

ºC, ºK).

SYSTEM I/O – allows set up and assigning of analyzer

Analog and Digital I/O.

SYSTEM – displays analyzer information (serial number,

Fat date, password, software version, launch/detect unit

temperatures), allows setting of date/time, gas type.

VALVE CONTROL ¬– allows for manual and/or automatic

control of the valve driver output signals.

SIGNAL PROCESSING – Factory set parameters only

LASER SPECTRA & CONTROL – displays spectra and

allows manual control of laser.

SYSTEM I/O - ANALOG OUTPUT

CHANNEL 1 to 3 – configuring each 4 to 20mA channel

to output Concentration, Transmission, Gas Temperature,

Gas Pressure or None.

WARNING MODE – setting of mA output response during

analyzer warnings (Block, Track, Hold).

FAULT MODE – setting of mA output response during

analyzer warnings (Block, Track, Hold).

BLOCK MODE LEVEL – specifies the analog output level

in block mode (high 20mA or low 3.3mA).

FIELD LOOP CHECK – allows specified 4-20mA output

levels to check and distinguish between the three analog

output connections; select analog output channel 1, 2, or

3 to check and input new value to output and press enter

to activate.

AO CH CALIBRATION – Pre-Calibrated at factory and not

normally required. Allows calibration of 4 to 20mA output

channels; follow onscreen instructions.