<7 ROUTINE MAINTENANCE> 7-8

IM 11Y01B02-01E-A 4th Edition September 11, 2012-00

THE WINDOW COULD BE HOT FROM THE HEATED CELL

 

 

• This process wetted window is constructed of sapphire and is therefore more durable than the

mirror surface. Use a clean, dry instrument air or nitrogen pressure (or canned air) supply to first

blow off any particulate matter.

• Using warm water and mild soap detergent, gently clean the window surface with a soft,

non-abrasive cloth.

• If the deposits do not come off then use a small amount of IPA and a soft, non-abrasive cloth.

• Use the same clean, dry instrument air or nitrogen pressure supply to blow dry the surface.

• Carefully check the entire surface of the window from different angles to ensure it is thoroughly

cleaned and ready for service.

• If the window does not appear to clean up well, then replace the window assembly with a new one.

• If the window appears to have an etched (or pitted) surface then it has probably been

contaminated with HF or other similar corrosive gas! A new window should be installed.

• Check and replace if necessary the window o-ring.