<5 INSTALLATION AND WIRING> 5-1
IM 11Y01B02-01E-A 4th Edition September 11, 2012-00
5 Installation and Wiring
5.1 Mounting of the Analyzer
Refer to the SAFETY section (Safe handling and relocation) for instructions on holding and moving the
analyzer.
The TDLS220 is suitable for wall mounting by means of the four corner located mounting holes as shown
below: The holed are ½” sized for typically 3/8” or ¼” bolts and should be used with appropriate sized
flat and locking washers. Ensure the analyzer is securing fastened and that there is suitable access for
maintenance, etc.
The analyzer is designed for operation outside buildings, at normal environmental conditions (see
Environmental specifications, page 11). However, it must be protected from direct rain and snow fall.
Mounting in a vibration free environment will ensure prolonged service life. Mounting in vibration prone
areas may introduce operational issues. Please consider mounting location with care.
The actual measurement is NOT affected by vibration, it is an instrument life-time reliability
consideration.
Figure 8 - Mounting Dimensions
22”
22”
22”