<3 GENERL SPECIFICATIONS> 3-2

IM 11Y01B02-01E-A 4th Edition September 11, 2012-00

Performance specification

Precision*

0.01% 0

2

Linearity*

Typically R2 > 0.999

Response time

5 or 10 seconds plus transport time to

analyzer

Drift*

Span drift (6-12 month calibration) <+/-

0.1% 0

2

Zero drift (6-12 month Calibration) <+/-

0.05% 0

2

Analog I/O (Optional) Outputs: Concentration/Transmission (3@

4-20mA isolated) Sub 4mA for warnings/
faults

Inputs: Pressure/Temperature Feed for

Compensation (2@ 4-20mA isolated,
powered or loop power)

Digital I/O

Outputs: Warning/Fault/Concentration Limit

Relays (3 Form C Relay SPDT rated 1A@
24VDC)
• Valve Control (3@ 24VDC, Max 10W per

valve), zero/span

• Inputs: Remote Validation (3 voltage free

floating contacts) for zero/span

Communications

Ethernet, IEEE 802.3, 10/100 Mbps, RJ45

Automatic USB data transfer (upload/
download settings and data)

Calibration

Recommended Calibration Check Interval

3-6 Months

Gas Sampling

The extracted sample should be typically

Conditions

filtered, clean and dry (non-condensing)

Cell volume = 260 cc

Flow rate of 1~20L/min, typically 6 L/min

Pressure of -3 psig to 100 psig

Temperature of -20°C(4°F) to 50°C (122°F

un-heated or 120˚C (248˚F) heated

Gas Measured* O

2

: 0.01% detection limit, Min Range

0-1%, Max range 25%

Performance Specifications are application dependant.

*Consult Yokogawa for ranges; All

performance specifications are for 25ºC at

1 bar.

Installation Specifications

By Design:

The analyzer is designed for operation

in General Purpose area. The addition

of a Purge System enables operation in

Hazardous Area for gaseous releases.

Class 1 Division 2 Group B, C and D

(ATEX/CE Pending) (Optional) CSA

Special Acceptance certification.

Flow Cell Wetted Parts

Standard: 316L, Sapphire windows,

Teflon encapsulated Viton O-rings, and

protected gold mirror

Optional: Monel Alloy 400, Kalrez 4079

O-rings

Mains voltage:

CE/230V model - 210 to 240V AC, 50-60

Hz, single phase 110V model - 100 to

130V AC, 50-60 Hz, single phase

Surge and temporary overvoltage

immunity in compliance with

EN 61326-1.

Integration

Configuration

• Sample is fully extracted from process (and should be

conditioned before measurement)

• Process pressure and temperature can be controlled or

the analyzer may require pressure and temperature inputs
(application dependant)

• Length of flow cell is fixed

Purge Gas & Validation

Available systems (standard or custom) for:
• Manual or Automatic Validation
• Manual or Automatic Calibration
• Manual or Automatic Stream Switch
• Analyzer purge gas control
• Other options Available

Display and Software Functions:

TruePeak software has multiple levels, the default (or start page) is

the Main Menu:

Main Menu Displays:

Gas Concentration

Transmission %

Status (warm-up, OK, Warning, Fault, etc.)

Temperature (Fixed, Active Ambient or Active)

Pressure (Fixed or Active)

Main Menu:

Basic Menu

Configure, 3 functions

View Spectra, 2 functions

Data, 3 sub-menus

Trends

Advanced Menu:

Configure, 9 sub-menus (User Password)

Calibrate & Validate, 3 sub-menus

Data, 4 sub-menus

Trends

Active Alarms:

List of active alarms

Shut Down Analyzer:

Instructions to close TruePeak local or VAC