<7 ROUTINE MAINTENANCE> 7-11

IM 11Y01B02-01E-A 4th Edition September 11, 2012-00

7.4 Validation and Calibration

There are several methods that can be used to validate and/or calibrate the TDLS220 TDL analyzer.

Generally, we recommend routine validation of the analyzer either on-line (if appropriately set-up) or off-

line.

Actual calibration should only be performed if certain performance criteria have not been met during the

validations and should only be performed by appropriately qualified personnel.

The options for Validation and Calibration are:

7.4.1 Off-Line Manual/Automatic Checking/Validation

Validate Off-Line (or Cal Check)

Manual (Zero - Span)

Manual introduction of zero or span gas and follow manual

procedure via user interface

Calibrate Off-Line

Manual (Zero – Span)

Manual or automatic introduction of Zero and Span gases for

validation or calibration (when equipped with appropriate hard-

ware, valves, etc.)

MANUAL OFFLINE VALIDATIONS

Enter into the Advanced Menu, Calibrate &

Validate section, Off-Line Validations.

Choose Check gas 1, 2 or 3 and select

Manual Validation.