<6 BASIC OPERATION> 6-12

IM 11Y01B02-01E-A 4th Edition September 11, 2012-00

ADVANCED CALIBRATE & VALIDATE MENU

OFFLINE CALIBRATIONS – allows zero calibration,

zero offset, span calibration, transmission adjustment,

dark current calibration, and peak search.

OFFLINE VALIDATIONS – allows manual or automatic

configuration of check gases 1 through 3.

ONLINE VALIDATIONS – DO NOT USE THIS

FEATURE on TDLS220

OFFLINE CALIBRATIONS – use for all TDLS220

ZERO CALIBRATION – manual or automatic

calibration of Zero.

ZERO OFFSET – allows manual adjustment of Zero by

applying a concentration offset.

SPAN CALIBRATION – manual or automatic

calibration of Span.

TRANSMISSION – adjustment of transmission value.

DARK CURRENT – calibration of detector dark

current.

PEAK SEARCH – initiates a system scan of absorption

peaks to validate current peak selection is correct.

OFFLINE VALIDATIONS

CHECK GAS 1 to 3 – allows manual or automatic

configuration up to 3 check gasses.

See section 5.5.4.