<6 BASIC OPERATION> 6-11

IM 11Y01B02-01E-A 4th Edition September 11, 2012-00

VALVE CONTROL

Typically used when the analyzer is configured with a

flow cell in an offline application.

VALVE CONTROL

The valve control includes manual and automatic

control.

MANUAL – allows for manual ON/OFF control of the

valve driver output signals.

TIME SEQUENCE – allows automatic valve switching

based on time sequence. It is normally used for

measurement that needs stream switching.

REMOTE OVERRIDE – allows remote valve control by

digital contact.

LASER SPECTRA & CONTROL

Displays Raw Detector Signal and Absorption

Spectrum as well as Gas Concentration, Gas

Temperature, Gas Pressure, Transmission, Laser

Temperature and Peak Center Position.

CAPTURE – allows a manual spectra capture (user will

be prompted to enter a unique file name for captured

spectra).

For Display Only:

CONTROL MODE – allows selection of Automatic

(laser temperature is controlled to keep peak centered

using peak center position as set point) or Manual

(laser temperature is controlled using integral laser

temperature sensor) – ONLY ADJUST WITH FACTORY

ASSISTANCE.

LASER TEMP – In manual mode allows adjustment of

laser temperature – ONLY ADJUST WITH FACTORY

ASSISTANCE.

LTSP LIMITS – setting of guard limits for laser

temperature set point – ONLY ADJUST WITH

FACTORY ASSISTANCE.

CUR CENTER – setting of center point for laser

current ramp –ONLY ADJUST WITH FACTORY

ASSISTANCE.