beautypg.com

9 data files and format> 9-4 – Yokogawa Extractive Oxygen Gas Analyzer TDLS220 User Manual

Page 76

background image

<9 DATA FILES AND FORMAT> 9-4

IM 11Y01B02-01E-A 4th Edition September 11, 2012-00

system

.bak

System History Back-Up

A back-up of historical log of

configuration changes

System Configuration History - When the system.his file exceeds

100KB size the contents is saved to this .bak file and the .his file

is emptied. This .bak file is in the form of ASCII data files that can

be opened with Microsoft “Notepad” as simple .txt file formats.

The content can then be copied and pasted into Microsoft Excel

spread. Each configuration change is logged with the date (MM-

DD-YYYY) and time that it occurred, the parameter (e.g. opl) and

the new value. Also, whenever the analyzer is started-up or shut-

down the ‘TruePeak Open/Close” parameter will be logged.

Example:

11-20-2006 08:13:10 opl(inch) 39.96 43.31

11-20-2006 08:13:26 temperature(F) 509.0 86.0

11-20-2006 08:13:40 pressure(psi) 14.50 14.50

system

.usr

System User

System configuration as shipped from

the factory

This contains all ‘Factory’ configuration settings as the analyzer

was shipped. This is used for reference purposes only.

valspe

.his

Validation Spectra

All spectra captured during On-Line and

Off-Line Validations

Spectra captured for historical Validations (On-Line and Off-Line)

are stored in this file.

valspe

.bak

Validation Spectra Back-Up

Backed-up spectra captured during On-

Line and Off-Line Validations

When the valspe.his file exceeds 1MB size the contents is

saved to this .bak file and the .his file is emptied. This .bak file is

Spectra captured for historical Validations (On-Line and Off-Line)

are stored in this file.

zero00

.spe

Zero Calibration Spectra

Absorption Spectrum and coefficients at

time of calibration

This file is essential to the analyzer calibration - if this file does

not exist or is corrupted, modified or otherwise tampered with

then the analyzer calibration is invalid. It contains essential

information relating to the Zero Calibration of the analyzer.

zero01

to

zero10

..spe

Historical Zero Calibration

Spectra

Previous Absorption Spectrum

and coefficients at time of previous

calibrations

zero01 through zero09 are previous files with zero01 being the

most recent. The most recent previous zero01 can be restored

in the analyzer as 'Previous Calibration'. The factory calibration

which can also be restored in the analyzer is named zero10.

zeroref

.cfg

System Configuration

At the time of current Zero Calibration

System Configuration at the time of current Zero Calibration

(zero00.spe)

memory

.res

File Size Management

for Results FACTORY PERSONNEL ONLY - used by TruePeak software to

manage the .res files

memory

.spe

File Size Management

for Spectra FACTORY PERSONNEL ONLY - used by TruePeak software to

manage the .spe files

This manual is related to the following products: