IM 11Y01B02-01E-A 4th Edition September 11, 2012-00

6.4 Stream Switching & Valve Control Outputs

The TDLS220 has three valve drivers (24VDC each @ 12W max each) terminated at

TB-3. These can be used for either calibration/Validation and/or stream switching

functions. If using these valve drivers for any type of Calibration/Validation function

then please refer to the appropriate section of Instruction Manual. As shown in the fol-

lowing drawing, 3 process streams are connected to 3 gas valves then to the inlet of

the TDLS220 flow cell. The gas valves are controlled by three 24V relay output signals

from the launch unit of the analyzer. During normal operation, only one valve can be

open at a time. The connected solenoid wires at TB-3 may require the installation of a

ferrite coil to eliminate noise spikes – this is totally dependant on the site specific elec-

trical grounding conditions. If the analyzer appears to be re-booting when one or more

of the solenoids is activated then this means some noise protection is required. Either

add a ferrite coil to the connecting wires (contact Yokogawa for further assistance if

necessary) or add a 3A 50V directional diode across the +/- terminals at TB-3.

Valve 1

Valve 2

Valve 3

Stream 1

Stream 2

Stream 3

Inlet

Outlet

TDLS220 Sample Flow Cell

24V relay control signals
from Control Unit

 

Figure 15 - Valve Control

Figure 15- Typical Stream Switching Scheme