<7 ROUTINE MAINTENANCE> 7-12

IM 11Y01B02-01E-A 4th Edition September 11, 2012-00

Following the on screen directions, enter in the:

Pressure= Cell Pressure

Temperature= Cell Temperature

Cell length= 40”

Concentration of the target gas inside the gas

flow cell. Press Enter to proceed.

Wait for the reading to stabilize. Press 9 to

proceed.

AUTOMATIC OFFLINE VALIDATIONS

The Offline Validations can also be

automatically configured. Refer to section

5.5.3 for details.

Local Initiate will start the automatic online

validation sequence when selected. It will use

the existing ‘Settings’ (see below for details on

‘Settings’).

Remote Initiate will enable/disable monitoring

of the selected “Remote Initiate” contacts on

TB-2 within the Launch Unit. When enabled,

the analyzer will detect the chosen contact

closure and automatically start the online

validation sequence.

Time Initiate will allow input of a specified

time to automatically start online validation

sequence once every day, every week, every 2

weeks, or every 4 weeks.