6.5.6 OPTIONAL Virtual Analyzer Controller (VAC) Operating Software Guide
Once an Ethernet connection has been established, the
VAC software can be started. To load the software on to
a laptop or desktop PC, follow these steps:
Create folder under c:\asiini
Copy all files from the CDROM to this folder
Create a shortcut from the desktop by browsing for the
ASI-VAC.exe file
Test the shortcut by double clicking on it and the VAC
software should open
Once the software is open, follow the on-screen options.
Function keys have been pre-assigned to certain tasks
allowing a RIU keypad to perform all necessary tasks.
The primary functions of the software are:
Create a virtual network computing connection to an
analyzer thus allowing for control of the analyzer, typically
for start-up, service, calibration, etc.
Allow for file transfer from an analyzer to a local USB port
(for memory device)
Create connections by name and/or IP address
OFFLINE CALIBRATIONS – use for all TDLS220
ZERO CALIBRATION – manual or automatic calibration of
Zero.
ZERO OFFSET – allows manual adjustment of Zero by
applying a concentration offset.
SPAN CALIBRATION – manual or automatic calibration of
Span.
TRANSMISSION – adjustment of transmission value.
DARK CURRENT – calibration of detector dark current.
PEAK SEARCH – initiates a system scan of absorption
peaks to validate current peak selection is correct.
After selecting “Copy Data – F3” you will be allowed to
select the analyzer (with description and IP address) you
wish receive data from.
This will initiate a data transfer for all results and
configuration files stored on the analyzer.