<6 BASIC OPERATION> 6-22

IM 11Y01B02-01E-A 4th Edition September 11, 2012-00

6.5.6 OPTIONAL Virtual Analyzer Controller (VAC) Operating Software Guide

Once an Ethernet connection has been established, the

VAC software can be started. To load the software on to

a laptop or desktop PC, follow these steps:

Create folder under c:\asiini

Copy all files from the CDROM to this folder

Create a shortcut from the desktop by browsing for the

ASI-VAC.exe file

Test the shortcut by double clicking on it and the VAC

software should open

Once the software is open, follow the on-screen options.

Function keys have been pre-assigned to certain tasks

allowing a RIU keypad to perform all necessary tasks.

The primary functions of the software are:

Create a virtual network computing connection to an

analyzer thus allowing for control of the analyzer, typically

for start-up, service, calibration, etc.

Allow for file transfer from an analyzer to a local USB port

(for memory device)

Create connections by name and/or IP address

OFFLINE CALIBRATIONS – use for all TDLS220

ZERO CALIBRATION – manual or automatic calibration of

Zero.

ZERO OFFSET – allows manual adjustment of Zero by

applying a concentration offset.

SPAN CALIBRATION – manual or automatic calibration of

Span.

TRANSMISSION – adjustment of transmission value.

DARK CURRENT – calibration of detector dark current.

PEAK SEARCH – initiates a system scan of absorption

peaks to validate current peak selection is correct.

After selecting “Copy Data – F3” you will be allowed to

select the analyzer (with description and IP address) you

wish receive data from.

This will initiate a data transfer for all results and

configuration files stored on the analyzer.