beautypg.com

4 analyzer components> 4-3 – Yokogawa Extractive Oxygen Gas Analyzer TDLS220 User Manual

Page 18

background image

FPGA
TTL out

Relay Coil
Drive Circuit

Relay Coil

Relay
Contacts
24VDC to
C and to NO

24VDC

TB3

DGND

24VDC 12W max to
external solenoid valve
when relay is energized

NOTE; Use ferrite coil or
direction diode on TB6
wired outputs to prevent
switching spikes

<4 ANALYZER COMPONENTS> 4-3

IM 11Y01B02-01E-A 4th Edition September 11, 2012-00

Calibration Valve Relays

There are three calibration valve relay circuits on the board. These are capable of actuating Form C SPDT

relays. The common pole is connected to 24VDC power and the normally open pole is routed to the field

terminal block. Digital ground is also routed to the terminal block TB3 as shown below.

Figure 3 - Calibration Valve Relay Diagram

Connections of each relay (Common and Normally Open) are routed through the board to field terminals.

The contacts are rated for a maximum of 1A @ 24VDC (or 0.5A @ 125VAC).

The pluggable field terminals are mounted on the lower edge of the board, just to the left side of the DC

power input terminals.

The appropriate relay(s) is actuated when a calibration gas check valve is to be initiated.

Laser Temperature & Current Control

The board has two main laser control function circuits, temperature control and laser current control.

Board Temperature

The board has a temperature sensing chip/circuit that monitors temperature of the board inside the main

electronics enclosure. The sensor is located on the top edge of the Back Plane.

Analog I/O Board outputs the analyzer results and reads input process gas compensation values (pressure

and temperature). The board has power status LEDs as well as voltage test points for the input and output

channels.

• Output channels (three) are ranged 4-20mA. They can be assigned to measured values

Oxgen, Transmission or compensation signal re-transmission.

• Input Channels (two) are used by the analyzer to read active values for process gas

temperture and/or process gas pressure. These are application dependant and may or may

not be required inputs. There are two channels, one for temperature and one for pressure.

Each may be used to read 4-20mA signals that are isolated or to read and loop power (with

integral 24VDC) signals.

Optional Mini Display (4x20 VFD) mounts on the analyzer enclosure door. The display itself is an indus-

trial grade 4 line 20 character vacuum fluorescence display (VFD) that is self illuminating (i.e. no back light

required).

This manual is related to the following products: