<6 BASIC OPERATION> 6-9

IM 11Y01B02-01E-A 4th Edition September 11, 2012-00

SYSTEM I/O – ANALOG INPUT

Pre-calibrated at factory and not normally required.

Allows calibration of 4 to 20mA input channels; follow

onscreen instruction.

SYSTEM I/O - DIGITAL OUTPUT

Setting of Digital Output assignments (DO 1-3)

CHANNEL 1 WARNINGS – Setting of levels that will

trigger analyzer warning and subsequent DO.

CHANNEL 2 FAULTS – Setting of levels that will trigger

analyzer fault and subsequent DO.

CHANNEL 3 USER ALARM – Setting of either Con-

centration or Transmission level (high or low) that will

trigger analyzer warning and subsequent DO.

FIELD LOOP CHECK – allows wiring connection check

during field installation.

CHANNEL 1 WARNINGS menu allows setting of

various analyzer WARNING conditions. Warning

is an event that will reduce but not eliminate the

measurement integrity, it is an indication that

maintenance is required but the analyzer is still

operational.

DETECTOR SIGNAL LOW – level at which low

detector signal will trigger a warning – ONLY ADJUST

WITH FACTORY ASSISTANCE.

TRANSMISSION LOW – level at which low

transmission will trigger a warning – ONLY ADJUST

WITH FACTORY ASSISTANCE.

SPECTRUM NOISE HIGH – level at which excessive

spectrum noise will trigger a warning - – ONLY

ADJUST WITH FACTORY ASSISTANCE.

PROCESS PRESSURE OUT OF RANGE – upper and

lower levels at which process gas pressure reading will

trigger a warning.

PROCESS TEMPERATURE OUT OF RANGE – upper

and lower levels at which process gas temperature

reading will trigger a warning.