beautypg.com

4 analyzer components> 4-6 – Yokogawa Extractive Oxygen Gas Analyzer TDLS220 User Manual

Page 21

background image

<4 ANALYZER COMPONENTS> 4-6

IM 11Y01B02-01E-A 4th Edition September 11, 2012-00

RIU Interconnect to TDLS220 Control Unit(s)

When connecting just one analyzer to the RIU there are two twisted pair wires to consider , there are only

four wires to be terminated to make the 10/100 Ethernet connection

RIU Optional Ethernet Switch

If there is more than one analyzer connected to the RIU, then they are routed via an industrial Ethernet

switch. Up to four analyzers can be routed through one RIU switch. The switch is powered by 24VDC from

the back-plane and includes several status LEDs.

Analyzer

Integral SBC

TB6

Tx+ Tx- Rc+ Rc-

RIU

Integral SBC

RIU Switch or
Field Terminals

Tx Tx Rc Rc

Figure 5 - Connecting RIU to Analyzer(s)

 

Analyzer 1 
SBC 

Analyzer TB6 
 
Tx Tx   Rc Rc 

Ethernet 
Switch 

Feed‐through 
Board 
Tx Tx    Rc Rc 

Analyzer 2 
SBC 

Analyzer TB6 
 
Tx Tx   Rc Rc 

RIU 
SBC 

Feed‐
Through 
Board 
Tx Tx    Rc Rc 

Figure 6 - RIU Ethernet Switch

This manual is related to the following products: